OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Display and Analysis of Images
LSM 510 META
DuoScan
4-340
B 45-0021 e
03/06
Buttons for Table
functions
:
Copy Table
button: The table of intensity values over Lambda is
copied to the clipboard.
Show Table
button: The table of intensity values over Lambda is
displayed on the bottom left of the Image Display window.
Save Table
button: The table of intensity values over Lambda can be
stored as a text file.
The toolbar below the intensity Lambda diagram has the following functions:
ChS1
button / slider
Selection of a wavelength range (use the sliders) assigned with the
spectral channel ChS1. This assignment can be used for scanning (when
Apply to hardware
is pressed, see
Config Control
) and /or for
extracting a single or multichannel image from the Lambda Stack (when
Extract channels
is pressed). Clicking on the color box in the ChS
buttons opens the color selection box and allows a new color
assignment to be made for this channel.
C /
−
button
Add (+) or reduce (-) the number of channels displayed and available for
assignment. Up to 8 channels labeled ChS1, ChS2, ChS3 ... ChS8 are
available.
Extract channels
button
Generates a new image or multicolor images based on the settings
made in the ChS1, ChS2, ChS3 ... ChS8 buttons. Two or more Lambda
channels are binned to form the channels ChS1 ... 8. The generated
image is displayed in a new
Image Display
window and is no longer a
Lambda Stack of images.
Fig. 4-315
3D graph