OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Acquire Menu
LSM 510 META
DuoScan
4-120
B 45-0021 e
03/06
Different XY Spot
Bleach Position
:
If this check box is ticked , you can set a different XY position for spot
bleaching. This function is only available using the
Spot
scanning mode. Click on
the
Spot Select
button in the
Scan Control
window. A new image is produced
and two crosshairs appear in the image. The red crosshair marks the spot that
will be imaged. The green crosshair marks the spot that will be bleached. Move
the center of the crosshairs to the desired positions and perform bleaching.
Store settings of the Bleach Control
Proceed as follows to store the entire settings of the
Bleach Control
window:
•
Enter a name in the Settings list box and click on
Store
to store the settings.
If required, stored settings for the bleaching procedure can be reactivated quickly.
•
Open the Settings list box with a click on the arrow button and select the required name with a click
of the mouse.
•
Then click on the
Apply
button to activate these settings. The
Bleach
Control
window will be
updated accordingly.
Delete settings of the Bleach Control
Settings which are no longer required can be deleted.
•
Open the Settings list box and select the required name.
•
Click on the
Delete
button. This stored setting will be removed.
The bleaching procedure can also be activated via keys 1 to 4 of the Trigger Control.
•
Open the
Trigger in
list box with a click on the arrow button.
•
Select one of the trigger keys 1 to 4 (e.g.
Trigger4
).
It is also possible to trigger an out signal via trigger control.
4.5.7.3
Bleach Parameter Panel
The
Bleach Parameter
panel allows you to
determine how often the bleaching process shall
be performed, and to select the area for bleaching
in the scan image via the
Define Region
button.
Fig. 4-99
Bleach Parameter panel