OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Options Menu
LSM 510 META
DuoScan
4-222
B 45-0021 e
03/06
4.9.5.12
Scan Mean of ROIs
The
Mean of ROIs
tab permits the presetting of
the
Image Display
window for the optional
MeanROI
function (time series) to be changed
with regard to scaling and display mode of the
intensity time diagrams.
(1)
Diagram Scaling
The following settings are possible by activating
one of the option buttons:
−
Automatic diagram scaling
−
Fixed time range for diagram time scale
; input of the time range in seconds via input box
−
Fixed number of cycles for diagram time scale
; input of the time range in number of cycles via
input box
(2)
Initial diagram types
The following settings are possible by activating the relevant option button:
−
One diagram
−
Channels diagram
−
ROIs diagram
On activation of the
Black graphs
check box, the intensity profiles in the diagram are displayed in
black (monochrome).
(3) Live
image
(a)
Scan the whole image
If you activate this check box, the complete live image will be scanned; only the defined ROIs will be
scanned if the check box is deactivated.
(b)
Save the whole time series
If you activate this check box, the complete Time Series will be scanned; only the Mean of ROI series will
be scanned if the check box is deactivated.
Fig. 4-217
Scan Mean of ROIs tab