OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Maintain Menu
LSM 510 META
DuoScan
4-242
B 45-0021 e
03/06
4.10.4
Set Find
This function permits the preset parameters of the
Find
function (see
Scan Control
matched individually.
•
Click on the
Set Find
button in the
Maintain
subordinate toolbar of the
Main
menu.
−
The
Auto B&C Control
window appears on
the screen.
•
Change the settings for the
Upper Threshold
of Data Depth
and
Gain Correction
using the
relevant sliders.
The settings can be made individually for each
detection channel or for all channels together.
For experiments with increasing of fluorescence
over the time it's necessary to reduce the
Upper
Threshold of Data Depth
for the
Find
function.
Each slider should be used separately.
•
Click on the
C
button to set the value for the
Gain Correction
to 100 %.
•
If required, activate the
Set Amplifier Gain to 1.0
check box.
•
Click on the
Find
button to start sample scanning with the current settings.
•
Clicking on
Default
enables you to activate the default settings again.
•
If required, activate
Auto B&C for all tracks
checkbox for
Find
function.
4.10.5 Spline
This function permits calibration of the Scanner position signals. This is required for the use of spline
curves in the
Line
scanning mode (see
Line
, page 4-92f).
Fig. 4-242
Auto B&C Control window