LSM 510
OPERATION IN EXPERT MODE
LSM 510 META
Extension of the LSM 510 or LSM 510 META with LSM DuoScan
Carl Zeiss
03/06
B 45-0021 e
4-355
•
Now click on the
Ch2
button in the
Channel Settings
panel to optimize Channel 1. Optimization is
performed in the same way as for the single channel and can be monitored online in the relevant
separate image of the channel.
•
Then optimize the second channel by clicking on the relevant button (
Ch3
) in the
Channel Settings
panel of the
Scan Control
window.
Now effect image optimization as explained for the single-channel mode, but separately for each
channel.
•
Now click on the
xy
button of the
Display
toolbar.
−
The composite scan image of two channels is presented in a common window.
Image optimization can be effected much faster if you select a smaller frame, since less data
have to be processed.
4.15
Extension of the LSM 510 or LSM 510 META with LSM DuoScan
The LSM 510 and LSM 510 META can be equipped with a second scanner, the LSM
DuoScan
. For optimal
results the scan fields of the two scanners must be aligned. This procedure is described in chapter 4.8.4.1
The additional equipment results in different screen contents of the
Bleach control
menu (see chapter
Fig. 4-336
Scan Control and Image Display windows