Olympus IX71 Microscope Z-Axis Motor Drive Installation

Applied Scientific Instrumentation - https://asiimaging.com/docs/

Olympus IX71 Microscope Z-Axis Motor Drive
Installation

This procedure steps you through the installation and alignment of the ASI’s IX71 Z-axis motor drive
onto the Olympus IX71 microscope.

NOTE: On January 2007 Olympus changed the size of the
threaded end of their fine focus shaft from 3 mm to 4 mm.
Microscopes with serial numbers '7Axxxxxx' and newer, and
all 8-digit versions, identify the 4 mm version. The threaded
shaft diameter can also be directly measured after the left-
hand fine-focus knob has been removed. This installation
manual applies to both versions.

Typical diameter of threaded end of Olympus 4 mm fine focus shaft

ASI offers Z-Drives with drive shafts for both 3 mm and 4 mm versions

Содержание IX71

Страница 1: ...1 microscope NOTE On January 2007 Olympus changed the size of the threaded end of their ne focus shaft from 3 mm to 4 mm Microscopes with serial numbers 7Axxxxxx and newer and all 8 digit versions ide...

Страница 2: ...The following tools are required for this procedure 1 16 Allen wrench provided 5 64 Allen wrench provided 3 32 Allen wrench provided 7 64 Allen wrench provided The procedure has three parts Removing...

Страница 3: ...the inner left ne focus knob with Allen wrench Loosen set screw in knob if present Use Allen wrench to remove screw that secures the outer focus knob Unscrew inner left ne focus knob while holding the...

Страница 4: ...ding right hand knob in place Slide exible coupler over the threaded end of the left ne focus shaft while holding the right ne focus knob in place and tighten clamp with a 2 mm or 5 64 inch Allen wren...

Страница 5: ...ch Allen wrench to remove the horizontal adjustment screw Use the 7 64 inch Allen wrench to loosen the two vertical adjustment screws Slide the baseplate assembly completely under the microscope as sh...

Страница 6: ...nstall https asiimaging com docs Printed on 2022 01 10 05 45 Slide base plate under microscope and center beneath focus assembly Leave base plate clamp loose at this time Leave base plate clamp slight...

Страница 7: ...in the horizontal adjustment screw to hold the drive assembly in place using the 7 64 inch Allen wrench Leave the adjustment screws loose enough so that later the drive can slide up and down and the...

Страница 8: ...ed or the drive may slip Lightly tighten the vertical and horizontal adjustment screws as shown Since the motor is usually self aligning this should be a good position for the drive Check the alignmen...

Страница 9: ...is felt repeat the above steps Part 4 Installing the Motor Drive Cover Plate and Fine Focus Knob Install cover with screws as shown Screw on small focus knob as shown Locate the motor drive cover Pos...

Страница 10: ...nstall the large outer knob that was removed in Part One Replace the rubber boot if 3 present After screwing on small knob attach the large knob Drive and cover installed on the Olympus IX71 This comp...

Страница 11: ...2022 01 10 05 45 11 11 Olympus IX71 Microscope Z Axis Motor Drive Installation Applied Scienti c Instrumentation https asiimaging com docs...