OPERATION IN EXPERT MODE
LSM 510
DuoScan
Carl Zeiss
Display and Analysis of Images
LSM 510 META
DuoScan
4-264
B 45-0021 e
03/06
4.13.10
Select - Reuse
This function allows to
−
transfer acquisition parameters of an image from the image data base to the
Microscope control
,
Configuration Control
,
Scan Control
,
Time Series Control
and
Bleach Control
windows and
applies those parameters directly on the system.
The acquisition parameters of an image are displayed in the
Image Display
window and can be viewed
by using the
Info
function. In the tab
Image Status Display
in the
Settings
function
of the
Options
subordinate toolbar it can be determined what parameters to view with the
Info
function.
The parameters include the following:
Frame Size, Speed, Data Depth, Scan Direction, Average, Zoom Rotation, Offset, Pinhole diameter,
Collimator, Detector Gain, Amplifier Offset, Amplifier Gain, Excitation, Beam Path and Scan Mode (Line,
Frame, Stack, Time Series).
However, the required objective must be selected by the user.
•
Click on the
Reuse
button. The acquisition parameters of the active image (stack) are applied
immediately to the system.
In the
Options
menu in the function
Settings
with the
Recording/Reuse
tab, it can be determined
whether the objective should also be transferred and set. Setting the microscope objective only works in
microscopes with motorized objective revolvers.