LSM 510
OPERATION IN EXPERT MODE
LSM 510 META
Purpose
Carl Zeiss
03/06
B 45-0021 e
4-11
4.1.3
Convention for the Text in this Manual
All the originally used terms of the software interface, e.g.
−
names of windows,
−
panels,
−
input boxes,
−
list / selection boxes,
−
check boxes,
−
menu items,
−
names of buttons and
−
keyboard keys,
are displayed in
bold letters
to allow easier identification.
4.1.4 Backup
System backup
−
A complete backup is contained on the enclosed backup CD-ROM.
User files backup
The following user-generated files need to be included in a backup procedure (keep directory structure):
−
Image database files: *.mdb (but not system_configuration_*.mdb)
−
LSM Image files: *.lsm
−
Exported images: *.* (*.Tiff, *.LSM-Tiff, *.BMP, …)
−
Palette files: AIM \ Palette \ *.lut
−
Filter files: AIM \ Filter \ *.krn
−
Pinhole setting files: AIM \ PH*.pos
−
Log files: AIM \ *.log
The following files generated during the system integration should also be included in a backup
procedure:
−
Parameter file for pinhole setting: AIM \ *.set
−
Parameter file after pinhole adjustment: AIM \ *.adj
−
Scanner files: AIM \ bin \ *.bin
−
Microscope stand database: AIM \ database \ system_configuration_*.mdb
4.1.5 Software
Operation
The LSM 5 software can be operated using the mouse, the PC keyboard, or both.
The operation of the mouse and the keyboard is identical to that of the Microsoft ® WINDOWS
operating system and is therefore not dealt with in detail in this manual.
If required, see the Microsoft manual or online help for relevant information.