background image

Summary of Contents for SV 6

Page 1: ...SV 6 and SV 11 Stereomicroscopes Operating instructions...

Page 2: ...SV6 Stereomicroscope 1 Binocular tube 2 Stereo body 3 Objective 4 Stand 5 Illuminator SV 11 Stereomicroscope...

Page 3: ...or transmitted fight es 14 Stands and stages Microscope carrier with pinion box e ee nanesen 16 StANAS Lio 16 Specimen stages 1 es 18 Documentation Phototubes and TV cameratubes erme 21 Drawing tube f...

Page 4: ...principle has been used in the product line for the SV 6 and SV 11 Stereomicroscopes permitting instrument configurations to be optimally adapted to the type and size of the specimen and the document...

Page 5: ...ng distance The stemi carrier should be in the middle of the focusing range The distance between the mount of the objective and the specimen is approx 90 mm for the Achromat S 1 0x objective Push secu...

Page 6: ...zoom system Knob 3 for stepless magnification setting between 0 8x 5x has click stops at 0 8x 1x 1 2 1 6 2 2 5x 3 2x 4x 5x If an objective S 1 0x and eyepieces 10x are used a total magnification of B...

Page 7: ...ailable The different objectives are used in the following specific fields Achromat S 0 63x 45 50 63 Objective with extremely good field flattening They image large object fields with edge to edge sha...

Page 8: ...clockwise and then while observing the image slowly 10 direction until the image is in focus Write down your dioptric values This ensures that the image sharpness of identical object features is maint...

Page 9: ...15 45 50 85 5 reduces the viewing angle of the binocular tube 45 from 45 to 30 and brings the viewing tube nearer to the observer It is used to advantage in combination with a drawing tube for example...

Page 10: ...photograph the illumination must be set to produce fewer shadows With directed light the brightness decreases as the square of the distance the brightness difference between the object and the backgro...

Page 11: ...the microscope carrier The illuminator 10 is integrated in the microscope carrier at the factory For subsequent installation proceed as follows e Insert hood of illuminator 10 3 from below or above in...

Page 12: ...stand column and 2 self supporting adjustable light guides 6 The two angeled light exit apertures can be turned through approx 90 and fastened in position using screw 8 To attach e Attach the all roun...

Page 13: ...on of the field of view is ensured from zoom factor 1x upwards Slit ring illuminator 41 70 68 Fig 12 This illuminator produces a homogeneous shadow free illumination of the object center at a working...

Page 14: ...ure in place of the plate with a 40 mm dia aperture in the 84 mm dia opening 5 of the unit The transillumination unit for brightfield darkfield is mainly used to enhance the contrast of colorless low...

Page 15: ...polarized light Fig 16 we recommend using the rotatable stage 45 51 20 5 transmitted light attachment 45 51 36 and the high intensity illuminator with focusing attachment 45 51 45 Insert polarizer 45...

Page 16: ...b Turning the screw clockwise will result in stiffer motion turning counterclockwise in easier motion This may be necessary when additional equipment is used such as a TV or 35 mm camera Stands Stand...

Page 17: ...y unevenness of the table top SW 3 key The base plate comprises a securing ring two stage clips and a black white insert plate Retrofitting of insert plates in the opening for 84 mm dia plates an inse...

Page 18: ...9001 4 with horizontally mounted coaxial drive knobs for specimen movement in an area of 76 mm x 50 mm and a mounting frame with glass plate 5 or mounting frames included in the Axiovert product range...

Page 19: ...g adapter 45 60 06 with inserted photo eyepiece S PI 10x or S PL 12 5x or MC 80 microscope camera using the same adapter with inserted projection lens P 2 5x 45 60 21 Reflex camera body with T 2 adapt...

Page 20: ...as already been focused on the object e Remove binocular tube S from stereo body e Mount drawing tube S on the right of the microscope fasten and push in slider 1 e Mount binocular tube S either direc...

Page 21: ...The following method is also frequently applied for cleaning optical surfaces Moisten a stick with cotton wool wrapped around it with pure alcohol and remove dirt How to avoid fungus growth In an ext...

Page 22: ...er at 2 which enables the user to point to any specimen area in the field of view The brightness of the movable arrow is adjustable and can be adapted to the background of the object 26 SV 11 Discussi...

Page 23: ...se fine focusing knob 6 to set the movable part of the pinion box in such a way that it is flush with the fixed part Loosen star knob 12 Set pinion box to height reguired on column and lock Push secur...

Page 24: ...Wir beraten Sie gern Carl Zeiss Jena GmbH Zeiss Gruppe Unternehmensbereich Mikroskopie D 07740 Jena Telefon 0 36 41 64 29 36 Telefax 0 36 41 64 31 44 Anderungen im Interesse der technischen Weiterentw...

Reviews: