5-30
Troubleshooting the Analyzer
BIOHAZARD
All products or objects that come in contact with human or animal body fluids
should be handled, before and after cleaning, as if capable of transmitting
infectious diseases. Wear facial protection, gloves, and protective clothing.
The operator should follow the recommendations to prevent the transmission of
infectious agents in health-care settings as recommended for potentially
infectious specimens in
Protection of Laboratory Workers from Infectious
Disease Transmitted by Blood, Body Fluids, and Tissue
, 2nd edition; Approved
Guideline (1997) Document M29-A, National Committee for Clinical Laboratory
Standards (NCCLS). This document contains complete information on user
protection and it can be used as reference material for instructions on laboratory
safety.
To remove and replace the front shear valve face or the shear-valve rotor
In the maintenance procedure,
Cleaning the Shear Valve
•
Go to
Step 1
to take the shear valve apart.
•
Go to
Step 3
to reassemble the shear valve with the new part.
To check analyzer performance
•
At the Operations menu, select the
Startup
tab and verify background
counts.
•
Run a whole blood primer.
•
Run controls. If controls do not recover, calibrate the affected channel.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...