Status Line messages
7-21
Corrective Action
1
Select
Off
at the analyzer touchpad.
2
Wait about 1 minute and select
On
to restart the analyzer.
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Centering Collar Retract Denied
To prevent the collision of autosampler components, the system did not execute
the command to retract the autosampler centering collar.
Autosampler - Critical Error, Reset Autosampler
Critical autosampler error. A reset is required.
Corrective Action
1
Select
Off
at the analyzer touchpad.
2
Wait about 1 minute and select
On
to restart the analyzer.
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Diagnostic Mode Not In Effect
The analyzer sent a diagnostic command to the autosampler while the
autosampler was in operating mode.
Corrective Action
1
Open the
Autosampler
window of the Exerciser tab.
2
Select
Enter Diagnostic Mode
.
Autosampler - Eject Rack Failed
The autosampler failed to respond to the eject rack command and is not clearing
racks from the mixer and input and output shuttles.
Possible Cause
1
Command
from analytical
module not
received by
autosampler.
Corrective Action
a
At the analyzer touchpad, select
Off
.
b
Wait about 1 minute and select
On
to restart the
analyzer.
c
If the problem persists, call Siemens Service for
assistance.
2
All
communication
See
1a, b, c
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...