7-92
Status Line messages
If the analyzer is locked in the standby state, verify that the pressure and vacuum
readings are within range and adjust if necessary. Select Standby to exit.
Status BTRIEVE - Alarm Not Defined
An alarm sent from the analyzer is unknown to the Data Manager (not defined in
the Alarm Dictionary).
Corrective Action
See Alarm Dictionary.
Status BTRIEVE - Test Not Found
A test defined in the Control Dictionary has been erased from the Test
Dictionary.
Corrective Action
Enter test information in the Test Dictionary.
Syntax Error in AlarmExc.par
The parameter file, AlarmExc.par, is not correctly defined, or there is a blank line
in the file. No exclusive analytical alarm will be taken into account in the patient
QC calculation.
Corrective Action
Call Siemens Service for assistance.
System Preparation Failed
An error occurred during the initial system preparation phase.
Possible Cause
1
Software
communication
failure
Corrective Action
Follow these steps to exit the ADVIA 2120/2120i
software and then restart.
a
Select
Shut Down ADVIA
in the Log On/Off
window.
b
Select
CTRL+ALT+DELETE
and log back onto the
system to restart the software.
2
Hardware error
Refer to associated hardware error message for
appropriate corrective action.
3
Faulty CPU
board
Call Siemens Service for assistance.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...