Regulatory Information
9-53
Performance Characteristics: Correlation Data
The performance of the ADVIA 120 Hematology System was compared with the
performance of the Technicon H
•
3 RTC
System (x) using whole blood from 60
normal donors and 70 hospital donors with each sample run in duplicate (n =
260).
The following regression statistics were computed using a protocol similar to that
recommended in the NCCLS document EP9-A,
Method Comparison and Bias
Estimation Using Patient Samples; Approved Guideline (1995)
.
Parameter Comparative
System
or Method
Regression
Equation (y=)
r Sy.x
%NEUT
Technicon H
•
3 RTC
1.02x - 0.6
0.997
1.6
%LYMPH
Technicon H
•
3 RTC
1.00x + 0.8
0.997
1.7
%MONO
Technicon H
•
3 RTC
0.85x - 0.3
0.943
0.8
%EOS
Technicon H
•
3 RTC
0.87x + 0.2
0.979
0.4
%BASO
Technicon H
•
3 RTC
0.67x + 0.0
0.772
0.2
%LUC
Technicon H
•
3 RTC
0.92x + 0.6
0.944
0.7
x = Technicon H
•
3 RTC system
y = ADVIA 120 Hematology system