
5-48
Troubleshooting the Analyzer
4.
Mark the tubes with tape (the top line is rinse and the bottom line is vacuum),
then remove the lines.
Be careful not to allow the tubes to fall behind the
aspirate paddle assembly. If necessary, use hemostats to hold the tubing
outside the assembly.
5.
Connect the lines to the new block.
6.
Install the block onto the aspirator assembly by aligning the tabs with the
appropriate slots. Push the block into the assembly and gently slide the block
upward until it clicks into place. Be careful that you do not pinch the wash or
vacuum line during the installation process.
7.
Check the length of the probe. (See page 5-3 for instructions.)
8.
Run several saline samples and verify that there are no leaks or bubbles in
the sample line.
9.
Run a whole blood primer to verify that the probe is being properly washed
and dried after aspiration.
10.
Run controls to verify analyzer performance.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...