8-12
Methods
Parameter Explanation
BASO % Saturation 100 x (BASO Saturation
÷
BASO PHA
Cells )
BASO Flatness
Sum of theSquared Differences
Mean Cell Counting Rate
9
×
LI
PMNx
÷
MNx
MNx
Peak X channel of Mononuclear cluster
MNy
Peak Y channel of Mononuclear cluster
PMNx
Peak X channel of Polymorphonuclear
cluster
Parameter Key
WBCB
White blood cell count from theBasophil/Lobularity method
# BASO
Absolute count of Basophils
% BASO
Percent of Basophils
% BASO Suspect
Percent of events from Baso Suspect area
% MN
Percent of events from Mononuclear area
% PMN
Percent of events from Polymorphonuclear area
% PMN Ratio
Ratio of the percent of the Polymorphonuclear events to the percent of NEUTS
and EOS obtainedfrom the peroxidase method
BASO % Dead Time
Percent of analysis time when the channel is busy and cannot detect flowcell
events
BASO % Noise
Percent of events from Noise area
BASO % Saturation
Percent of events from Saturation area
LI
Lobularity Index
% Blast
Percent of events from Blast area