
Troubleshooting the Analyzer
5-25
15.
Gently push plunger on the cleaning syringe to flush bleach solution through
the RBC flowcell
(
2
)
. Repeat with deionized water.
16.
Reattach the shuttle line
(
10
)
to the side of the UFC
(
11
)
.
17.
Refill the syringe with bleach solution.
18.
At the Valves window of the Exerciser tab, open V20 and 21 and then push
the syringe to push bleach solution through the shuttle pathway to the VSC
chamber. Repeat with deionized water and then close V20 and V21.
19.
Remove syringe
(
4
)
from the white fitting.
20.
Reattach the line to the fitting on V23
(
3
)
.
21.
Select the Analyzer
Status tab to exit the Exerciser.
22.
Open the
Hydraulic
Functions
tab and then perform one System Wash
cycle.
23.
Run controls to verify system performance.
Repair and Replacement
Replacing the Clot Filters
Materials required:
•
Lens tissue
•
Replacement clot filter,
PN 113-3719-01
Time: Replacement - 5 minutes
Checkout - 5 minutes
Analyzer mode: Standby
1.
Unscrew the fitting and tube assembly (
1
) from the front of the selector valve (
2
).
2.
Gently push the silicone sleeve (
3
) to release the clot filter (
4
) from the filter
adapter fitting (
5
).
3.
Visually check the filter adapter fitting and the input port of the selector
assembly for any dirt or small particles.
4.
If you find some debris, use a piece of lens tissue moistened with water to
remove it.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...