Status Line messages
7-39
This error can
occur during the
following
analytical cycles:
•
Adjust Gains -
RBC/RETIC
•
Align Optics
•
Direct Optics
Follow these steps to exit the ADVIA 2120/2120i
software and then restart:
1
Select
Shut Down ADVIA
at the Log On/Off
window.
2
Select
CTRL+ALT+DELETE
and log back onto the
system to restart the software.
If the problem persists, call Siemens Service for
assistance.
B
Bad Analysis for R Message :10402
The results message sent to the host is incorrect.
Corrective Action
Call Siemens Service for assistance.
Bad Analysis for Y Message :10402
The format of the received workorder is incorrect. The workorder cannot be
created.
Corrective Action
Call Siemens Service for assistance.
Bad Mixer Cal Value - Reset Autosampler
Mixer-aspirate calibration failed while the system was in diagnostic mode
(Exerciser).
The corrective actions for this error must be performed by Siemens Service
personnel only.
Possible Cause
1
Autosampler
Calibration
error
Corrective Action
Calibrate the mixer aspirate position again.
2
Autosampler
CPU board
fails to store
calibration data
in flash
memory
Replace autosampler CPU board.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...