Status Line messages
7-11
4
Clogged clot
filter.
Remove clot filter and discard as biohazardous material.
Replace clot filter.
5
Loose
connection in
sample line.
Check connections and tighten if necessary.
6
Clog in needle.
Perform a needle rinse. Select
Utilities
menu,
Hydraulic
Functions
tab,
Needle
Rinse
.
7
Bent or
damaged
needle.
Replace the needle.
Follow the steps in
Replacing
the
Sampler Needles
.
8
Faulty
Conductivity
Detector
Contact your Siemens Field Service representative for
assistance.
Aspiration Paused
The analyzer is unable to perform a sample aspiration or to start a hydraulic cycle
when certain menus or tabs are open or when certain error messages are
triggered. Also the green ready-to-aspirate light is off.
Possible Cause
1
Aspiration is paused while
the following menus or tabs
are selected:
Reagent
Installation
,
Backup/Restore
,
System
Setup
,
Tools Modify
, or
with the triggering of some
error messages.
Corrective Action
Select any tab or menu button other than
the ones identified at left under Possible
Cause.
2
Critical system error
occurred, transferring the
current system state to
"Aspiration Paused."
Refer to the associated error message for
appropriate corrective action steps.
Atypical Lymphocytes - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...