7-22
Status Line messages
between the
autosampler
and the
analyzer may
be down.
3
Autosampler
operation may
be down.
See
1a, b, c
Autosampler Error - Reason Unknown
Unknown autosampler software error.
Corrective Action
1
At the analyzer touchpad, select
Off
.
2
Wait about 1 minute and select
On
to restart the analyzer.
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Error, Reset Autosampler
The autosampler has experienced a software error and must be reset.
Corrective Action
1
At the analyzer touchpad, select
Off
.
2
Wait about 1 minute and pr select
On
to restart the analyzer.
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Extra Rack Ejected
This message is for your information only. If the message persists, call Siemens
Service for assistance.
Autosampler Failure - Stop
There has been an autosampler communication error. A reset is required.
Corrective Action
1
At the analyzer touchpad, select
Off
.
2
Wait about 1 minute and select
On
to restart the analyzer.
If the problem persists, call Siemens Service for assistance.