7-38
Status Line messages
Autosampler - Startup Failed, Reset Autosampler
The autosampler has failed to start up. A reset is required.
Corrective Action
1
Open the autosampler access door.
2
Close the autosampler access door. The autosampler resets.
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Tube Too High
Tube is sitting too high in rack, or tube exceeds specifications, halting
autosampler.
Possible Cause
1
Tube is
improperly
positioned.
Corrective Action
a
Push tube down.
b
At the analyzer touchpad, select
Start/Stop
to
continue operations.
2
Tube is too
long.
a
Remove tube from rack and aspirate manually.
b
At the analyzer touchpad, select
Start/Stop
to
continue operations.
Autosampler Startup Complete
The autosampler has successfully completed the startup process.
Autosampler - Unexpected Car Position, Reset Autosampler
The autosampler car is not in the proper position. A reset is required.
Corrective Action
1
Open the autosampler access door.
2
Close the autosampler access door. The autosampler resets.
3
If the problem persists, call Siemens Service for assistance.
Awaiting Header
Software communication error.
Possible Cause
Corrective Action
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...