Status Line messages
7-53
The parameters in Customer Parameters §21, ASSOCIATE CBC/DIFF IMAGE
WITH A TEST GROUP and ASSOCIATE RETIC IMAGE WITH A TEST
GROUP, can be set only to values attached to a test group. The test groups are
defined in Customer Parameters §17.
Corrective Action
1
Open
Customer Parameters
in Tools Modify mode.
2
Define the test groups in Customer Parameters §17.
3
Set the values in §21 to correspond to the test groups you have defined.
or
1
Open
Customer Parameters
in Tools Modify mode.
2
Set the values in §21 =
0
.
Immature Granulocytes - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
Immature Granulocytes - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The autosampler halts at the same time the stop
message occurs. The specified stop criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria. The counter associated with the Stop
Criteria is reset when the Start/Stop button is selected.
Indicator and Switch Error - Reset Required
Potential software communication error or a hardware failure associated with the
front panel switch/indicator.
Possible Cause
1
Hardware failure
Corrective Action
a
Reset the analyzer by selecting
Utilities,
Exerciser, Indicators, Analyzer Reset
.
b
After about 1 minute, if the status line does
not display Ready to Run and the green
ready-to-aspirate light is still off, follow
these steps to exit the ADVIA 2120/2120i
software, turn off the analyzer and then
restart.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...