6-36
Flags
•
WBCB and WBCP agree within specified limits (No WBC-CE flag)
•
[(%NEUT + %EOS) - %PMN] is between 0 and 7.5
Results Flagged
WBCP, %NEUT, #NEUT, %LYMPH, #LYMPH, %MONO, #MONO, %EOS,
#EOS, %LUC, and #LUC
Corrective Action
Multiple occurrences of this flag, especially for consecutive samples, can indicate
a system problem. However, isolated instances of this flag are usually sample
related.
The following list of sample-related causes may not contain all conditions that
could cause this flag. There is no intention to associate the flag with specific
diagnoses.
System Related
Sample Related
1.
Check perox reagents.
2.
Check peroxidase hydraulics and sample
delivery.
3.
Check delivery of SHEATH/RINSE.
4.
Check PEROX SHEATH delivery.
5.
Check the pressure and vacuum readings.
6.
Check the perox reaction chamber
temperature.
7.
Check perox flowcell alignment.
8.
Check perox gains.
•
Neonatal samples
•
Aged blood sample
•
Malaria parasites
Perox Noise (PX-NO, NX)
Definition
The Perox Noise flag is triggered if 60% or more of the perox signals come from
the Noise area of the Perox cytogram.
This flag is triggered if 60% or more of the PEROX events are counted in the
Noise area.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...