5-26
Troubleshooting the Analyzer
5.
Wet the new clot filter with water and place it into the filter adapter fitting.
Orientation of the filter is not important.
6.
Reconnect the filter adapter fitting to the selector valve. Finger tighten.
IMPORTANT
Be very careful not to misthread the fitting. Misthreading could strip the
valve which will prevent proper operation.
7.
Run a sample and visually check for air bubbles in the sample line between
the selector valve and the UFC block.
8.
Check aspiration rate.
9.
If air bubbles appear, check that the filter adapter fitting is screwed on
correctly and that it is finger-tight, then repeat step
7
. If air bubbles still
appear, disconnect the filter adapter fitting and check for dust or debris in the
fitting and in the port.
If air bubbles still persist, call Siemens technical support.
Replacing the Diaphragm Pumps
Time: Replacement - 5 minutes
Checkout - 15 minutes
Materials required:
1
DP1 - PEROX SHEATH
2
DP2 - SHEATH /RINSE
3
DP3 - SHEATH /RINSE
4
DP4 – WASH
PN
067-B294-04
5
DP5 - SHEATH /RINSE
PN
067-B294-01
Analyzer mode: Off
REMOVE BOTTLES TO ACCESS DIAPHRAGM PUMPS
1.
Remove the reagent bottles from the side where you are replacing the
diaphragm pump.
2.
Pull the pump forward until it is released from its keyhole slots.
3.
Remove the three tubes, then attach these tubes to the new pump.
4.
Insert the raised studs on the top of the pump into the keyholes and push the
pump back until it is secure.
5.
Replace the reagent bottles.
6.
Checkout
For all replaced pumps - prime reagent lines and verify that there are no leaks.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...