5-22
Troubleshooting the Analyzer
The operator should follow the recommendations to prevent the transmission of
infectious agents in health-care settings as recommended for potentially
infectious specimens in
Protection of Laboratory Workers from Infectious
Disease Transmitted by Blood, Body Fluids, and Tissue
, 2nd edition; Approved
Guideline (1997) Document M29-A, National Committee for Clinical Laboratory
Standards (NCCLS). This document contains complete information on user
protection and it can be used as reference material for instructions on laboratory
safety.
CAUTION
When performing this procedure, do not use force on the syringe at any time.
1.
At the Utilities
menu, select
Exerciser
to open the Exerciser tab.
2.
Disconnect the waste tubing
(
1
)
that comes from the Perox flowcell
(
2
)
at
V24. This is the white fitting at V24
(
3
)
.
3.
Fill the cleaning syringe
(
4
)
EZ KLEEN and attach it to
the white fitting. Refill the
syringe EZ KLEEN as needed
during the following steps.
4.
Disconnect the sample tubing
(
5
)
from the top of the sample
syringe
(
6
)
and place the
sample tubing into a beaker to
catch fluid as you backflush
the flowcell
5.
Gently push the plunger on
the cleaning syringe to flush
the EZ KLEEN through the
Perox flowcell
(
2
)
. Repeat
with deionized water.
6.
Reattach the sample tubing
(
5
)
to the top of the sample
syringe
(
6
)
V15
V8
V5
V6
V7
V16
V17
V18
V25
V24
1
2
3
4
5
7
8
7.
Refill the syringe with EZ KLEEN.
8.
Disconnect the shuttle line
(
7
)
from the side of the UFC
(
8
)
and place the
sample tubing into a beaker to catch fluid as you backflush the flowcell.
9.
Gently push the plunger on the cleaning syringe to flush the EZ KLEEN
through the Perox flowcell
(
2
)
. Repeat with deionized water.
10.
Reattach the shuttle line
(
7
)
to the side of the UFC
(
8
)
.
11.
Refill the syringe with EZ KLEEN.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...