Regulatory Information
9-5
Do not assay cold specimens!
Allow the specimen to equilibrate to room temperature before mixing.
Do not
warm specimen by placing it into a 37°C water bath or incubator.
CAUTION:
If the whole blood specimen is brought to
room temperature quickly, there is a strong tendency for
the platelets to clump upon mixing.
Do not use previously frozen whole blood samples.
Freezing the blood can disrupt the cellular structure, thus
resulting in aberrant cell counts.
Unless otherwise stated, no special treatment of the whole blood specimen is
required. However, samples must be collected in the specified collection tube and
gently, but thoroughly mixed at the time of collection and again before sampling.
Calibrator and Control Products
CAUTION! POTENTIAL BIOHAZARD
Contains human source material. While each human serum or
plasma donor unit used in the manufacture of this product was
tested by FDA-approved methods and found nonreactive for
hepatitis B surface antigen (HBsAg), antibody to hepatitis C
(HCV), and antibody to HIV-1/2, all products manufactured using
human source material should be handled as potentially
infectious. Because no test method can offer complete assurance
that hepatitis B or C viruses, HIV, or other infectious agents are
absent, these products should be handled according to established
good laboratory practices.
Siemens supplies control and calibrator products for use with the ADVIA
2120/2120i Hematology System.
Refer to the current Price List, and the product package insert for additional
information.
ADVIA 120 TESTpoint 3-in 1 Hematology Control Abnormal 1
Product
Number
Symbol Contents
Amount
(mL)
09459683 T03-4417-54
Hematology Control
Abnormal 1
4 x 4.0 mL
- T03-4417-01
Hematology Control
Abnormal 1
4.0 mL
ADVIA 120 TEST Point 3-in-1Hematology Control Abnormal 2
Product
Number
Symbol Contents
Amount
(mL)
03410380 T03-4418-54
Hematology Control
Abnormal 2
4 x 4.0 mL
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...