Status Line messages
7-67
OPEN - Sid# Already Exists
The host has sent a workorder to the Data Manager that already exists in the
database. The Data Manager cannot update the current workorder with the new
one because the sample status is Incomplete, Current Complete, Complete, or All
Complete.
P
Parallel Node Error - Reset Required
Potential software communication error or hardware failure associated with the
Parallel Node Board.
Possible Cause
1
Software
communication
error or hardware
failure
Corrective Action
a
Reset the analyzer by selecting
Utilities,
Exerciser, Indicators, Analyzer Reset
.
b
After about 1 minute, if the status line does not
display Ready to Run and the green ready-to-
aspirate light is still off, follow these steps to exit
the ADVIA 2120/2120i software, turn off the
analyzer and then restart.
i
Select
Shut Down ADVIA
in the Log On/Off
window.
ii
Select
Off
at the analyzer touchpad.
iii
Select
CTRL+ALT+DELETE
and log back onto
the system to restart the software.
iv
Restart the analyzer by selecting
On
at the
analyzer touchpad.
2
Faulty power
supply, fuse,
cable, Parallel
Node Board, or
Can Bus
Scrambler Board
Call Siemens Service for assistance.
Patient QC Formula Not Defined
This error occurs when there is no value defined for the FSE parameter
Calculs_PatientQC_Formula. This parameter determines the formula used to
calculate the moving average. To define the parameter, contact your local
Siemens technical support provider.
The possible settings for the parameter are:
0 = parameter not active
1 = Bull’s Algorithm
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...