5-18
Troubleshooting the Analyzer
CAUTION
To avoid getting fingerprints on the glass windows of the flowcell, always
hold the flowcell by the metal slides.
Cleaning the Flowcells off the Analyzer: Step 2
Taking the Flowcell Apart
1.
Unscrew the locking nut (
1
) at the input end of
the flowcell assembly.
2.
Using a twisting motion, remove the CFM (
2
)
from the flowcell.
3.
Be careful not to lose the O-ring seal (
3
).
Cleaning the Flowcells off the Analyzer: Step 3
Cleaning the CFM
1.
A pin vise (
1
) and a 1/64-inch (0.4 mm)
diameter drill bit (
2
) are supplied in the
Flowcell Cleaning Kit. Open the pin vise, insert
the drill bit approximately 1/4 inch (6.4 mm)
into the vise jaws, then tighten the pin vise.
2.
Insert the drill bit into the sheath nipple (
3
).
Using a twisting motion, carefully push the
drill bit through the sheath nipple. If a clog is
encountered, proceed slowly while continuing
to twist the drill bit.
3.
Repeat step
2
for the sample input nipple (
4
).
4.
Carefully insert the bit into the shuttle nipple
(
5
) to a depth that just barely leaves the flutes
of the bit extended beyond the end of the
nipple. Twist the drill bit a few times, then
remove it from the nipple.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...