Status Line messages
7-31
1
Select
Off
at the analyzer touchpad.
2
Wait about 1 minute and select
On
to restart the analyzer.
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Needle Retract Denied
To prevent the collision of autosampler components, the system did not execute
the command to retract the autosampler needle.
Autosampler - No Tube Present
Pushpin cannot find tube in current rack position. Rack spring clip is broken and
holds tube incorrectly.
Corrective Action
Remove rack and check spring clip.
Autosampler - Operating Mode Not in Effect
The autosampler is not in operating mode.
Corrective Action
1
Open the
Autosampler
window of the Exerciser tab .
2
Exit the
Exerciser
tab by opening another tab.
Autosampler - Output Queue Full
Possible Cause
1
Output queue
and/or output
shuttle are full.
Corrective Action
Remove racks from output queue and/or output shuttle.
2
Output queue
sensor, ribbon
cable, or
autosampler
CPU board is
faulty.
CAUTION
Call Siemens Service for assistance. Only Siemens
Service personnel are authorized to take these corrective
actions:
a
Check output-queue sensor and sensor wiring.
b
Make sure pressing output-queue sensor flag triggers
sensor.
c
Make sure output-queue ribbon is plugged in
securely.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...