8-8
Methods
Baso Noise Threshold
This threshold (shown as a
red line) is set at channel 4
on the y axis.
Events below this threshold
are counted as noise, while
cells above the threshold are
included in the white cell
count (WBCB).
Basophil Threshold
This threshold (shown as a
red line) separates the
Basophils from the nuclei of
the other cell types.
The threshold is set at
channel 21 on the y axis. All
cells above this threshold
are counted as Basophils.
Baso Saturation Threshold
This threshold (shown as a
red line) is set at channel 45
on the x axis and channel
21 on the y axis.
Events within this area
(microbubbles, clumped
cells, and unlysed cells) are
monitored for flagging
purposes.