6-44
Flags
1.
Check retic reagents.
2.
Check RBC/Baso/Retic hydraulics.
3.
Check retic hydraulics.
4.
Check sheath delivery.
5.
Check the pressure and vacuum readings.
•
Aged blood
•
Transfusion
•
Sickle cell anemia
•
High numbers of
NRBCs
Retic Absorption Flatness (RTC-FL, RF)
Definition
The Retic Absorption Flatness flag is triggered when the CV (coefficient of
variation) for the RETIC ABS Flatness histogram is greater than 3.6.
Results Flagged
%RETIC, #RETIC, CHg, CHr, CHCMg, CHCMr, CHDWr, CHDWg, MCVg,
MCVr, RDWg, RDWr
Corrective Action
Laser oscillation (mode hopping)
Laser oscillation (also called mode hopping) is defined as a change in the laser
output frequency during the analysis. Typically, this condition triggers three
sample/system flags.
•
RETIC Fit Suspect flag (RTC-FS) is triggered due to the bimodal absorption
populations that appear on the RETIC Scatter Absorption cytogram (
1
) and
the RETIC Absorption histogram (
2
).
•
RETIC Absorption Distribution Abnormal flag (RTCADA) is due to the
bimodal absorption populations that appear on the RETIC Absorption
histogram (
2
).
•
RETIC Absorption Flatness flag (RTC-FL) is triggered due to the large
variation in absorption signals that appears on the RETIC ABS Flatness
histogram (
3
).
In such cases, the operator must rerun the sample.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...