7-40
Status Line messages
Bad Test in Profile
An incorrect test code is defined in a profile.
Corrective Action
Correct the Profile Table using the List menu in the Table Dictionaries window.
BASO Count Suspect - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
BASO Count Suspect - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The autosampler halts at the same time the stop
message occurs. The specified stop criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria. The counter associated with the Stop
Criteria is reset when the Start/Stop button is selected.
BASO Irregular Flow Rate - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
BASO Irregular Flow Rate - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The specified stop criterion is user definable in the
Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Stop Criteria is reset when the Start/Stop button is selected.
BASO Reagent Empty - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The reagent is empty, and the autosampler has
halted. The specified alarm criterion is user definable in the Customize Menu,
System Setup, Alarm/Stop Criteria, Reagent Conditions.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...