7-4
Status Line messages
Possible Cause
1
PSI pressure
gauge is out of
range
Corrective Action
Check the 20 PSI gauge and adjust the regulator if
necessary.
Setpoint: 20 PSI
±
1.
2
Pressure leak
Check for crimped or disconnected hydraulic or
pneumatic lines. Straighten or reconnect lines as
necessary.
20 PSI Out of Range - Stop
System pressure is out of range. The system has stopped because the pressure has
reached a critical level that may affect the integrity of sample results and damage
the system.
Possible Cause
1
PSI pressure
gauge is out of
range
Corrective Action
Check the 20 PSI gauge and adjust the regulator if
necessary.
Setpoint: 20 PSI
±
1.
2
Pressure leak
Check for crimped or disconnected hydraulic or
pneumatic lines. Straighten or reconnect lines as
necessary.
IMPORTANT
If problems persist, Call Siemens Service for assistance. Further corrective action
must be taken by Siemens service personnel only.
4
4 Key Not Found
The Control Dictionary or the Alarm Dictionary is empty. If the Alarm
Dictionary is empty, sample/system flags will appear as exclamation points (!).
Possible Cause
1
The Data Manager
received a sample that is
identified as a control but
that has not been set up in
the
Control Dictionary
.
Corrective Action
Do not use control ID labels on samples, or
set up the control in the
Control
Dictionary
.
2
The
Alarm Dictionary
is
empty or missing the Delta
Check alarm. The Delta
Check alarm must be in
the
Alarm Dictionary
even if you do not use
delta checking.
Enter alarms in the
Alarm Dictionary
.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...