Status Line messages
7-15
Autosampler - Autosampler is Already Running
The autosampler received a command to start or eject racks while it was running.
Autosampler - Bad Mixer Aspiration Cal
Mixer-aspiration calibration failed in diagnostic mode (Exerciser tab).
CAUTION
Call Siemens Service for assistance. Only Siemens Service personnel are
authorized to take these corrective actions.
Possible Cause
1
Autosampler
calibration
error.
Corrective Action
Calibrate mixer aspirate position again.
2
Autosampler
CPU board
failed to store
calibration data
in flash
memory.
Replace autosampler CPU board.
Autosampler - Bad Mixer Shuttle Cal
Mixer shuttle/index calibration failed in diagnostic mode (Exerciser tab).
CAUTION
Call Siemens Service for assistance. Only Siemens Service personnel are
authorized to take these corrective actions.
Possible Cause
1
Autosampler
calibration
error.
Corrective Action
Calibrate mixer shuttle/index position again.
2
Autosampler
CPU board
failed to store
calibration data
in flash
memory.
Replace autosampler CPU board.