8-68
Methods
Parameter Explanation
IRF-M+H
100 x ([# #MRetic]
÷
RETIC
Count)
RTC Mean X
Mean x channel of the RETIC Scatter
cytogram
RTC Mean Y
Mean y channel of the RETIC Scatter
cytogram
#RTC Cells Acquired
Number of Valid Signals Obtained
#RTC Cells Analyzed
Cells on RTC Scatter cytogram with
nonzero volume and hemoglobin
values, and are not in channel 99
#RTC Gated Cells
#Neg RBC + Retic Count within
threshold limits
Slope Negative Cells
Slope of the negative (mature RBC)
population relative to y axis of the
RETIC Scatter ABS cytogram
Parameter Key
#RETIC
Absolute number of reticulocytes
%RETIC
Percent of reticulocytes
CHCMr
Reticulocyte Cellular Hemoglobin Concentration Mean
CHr
Reticulocyte Cellular Hemoglobin Content
MCVm
Mean Cell Volume of mature population
MCVr
Mean Cell Volume of reticulocyte population
RBC
The system always uses the RBC count from the RBC / Platelet method, even
when a sample has been run with a retic only selectivity.
RTC RBC
Red blood cell count from the Reticulocyte method. The RTC RBC count is
calculated in the same manner as the RBC count from the RBC / Platelet Method.
RTC % Dead Time
Percent of analysis time when the channel is busy and cannot detect flowcell
events
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...