7-34
Status Line messages
Possible
Cause
1
Pushpin is
jammed.
Corrective Action
a
Open the autosampler access door.
b
Close the autosampler access door. The autosampler will
reset.
c
If problem persists, select
Off
at the analyzer touchpad.
d
Wait about 1 minute and select
On
to restart the analyzer.
e
If the problem persists, call Siemens Service for
assistance.
2
Pushpin
catches on
the bottom
of an
improperly
positioned
rack.
a
Check rack position in mixer. Adjust if necessary.
b
If the problem persists, call Siemens Service for
assistance.
3
Pushpin
catches on
the bottom
of the
mixer.
a
Check rack position in mixer. Adjust if necessary.
b
If the problem persists, call Siemens Service for
assistance.
4
Faulty
output-
queue
sensor or
wiring.
Call Siemens Service for assistance.
Autosampler - Pushpin Failed to Retract, Reset Autosampler
The autosampler pushpin failed to retract properly.
Possible Cause
1
Pushpin is
jammed.
Corrective Action
a
Open the autosampler access door.
b
Close the autosampler access door. The autosampler
will reset.
c
If problem persists, select
Off
at the analyzer
touchpad.
d
Wait about 1 minute and select
On
to restart the
analyzer.
e
If the problem persists, call Siemens Service for
assistance.
2
Pushpin
catches on the
a
Check rack position in mixer. Adjust if necessary.
b
If the problem persists, call Siemens Service for
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...