Daily Routine
3-9
Validating the Results
1.
At the Data Manager menu, select
Review / Edit
.
2.
If not already done, select validation mode and access mode to determine
which sample records will be reviewed.
3.
Review the displayed results. Scroll to view additional results.
4.
You can:
•
Go to step 5 if all results are acceptable.
•
Edit a result.
a.
Select
C.RES
(current result) box for test.
b.
Enter new value or comment code.
c.
Select
End
to exit, or
Next
to edit additional results.
You can select
Individual
to edit one result,
Successive
to access
all results, or
Pending
to access only missing results.
•
Apply a disposition to an individual result.
a.
Right-select the
D
column next to the test you want to receive the
disposition.
b.
Select the desired disposition.
None
cancels any previous disposition.
Rerun
repeats test with no dilution using the same sample ID. The
letter R displays in D box.
Exchange
swaps information in C.RES and PREV.RUN boxes. The
letter X displays in D box.
Accept
indicates result approved by operator. The letter A displays
in D box.
Delete
erases the result. The letter D displays in D box.
Dil.Sample
repeats test with specified dilution using a different
sample ID. When result is available, select
Dil.Cons
(Diluted sample
- Consolidation) to enter result into original sample record.
Only Accept and Delete are available for control samples.
The disposition takes effect when you validate the results.
•
Apply a disposition to all results.
a.
Select
Global Disposition
, or select the
Sample
menu and then
choose
Global
Disposition
.
b.
Select the desired disposition. (See available dispositions above.)
The disposition takes effect when you validate the results.
5.
Select
OK
to validate the sample results.
NOTE:
Be sure to complete sample validation (steps 1 through 5) before
rerunning a sample. If not, the sample is rerun with the default selectivity.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...