Troubleshooting the Analyzer
5-43
LASER WARNING
To avoid damage to the eyes, never look directly at the laser beam or at its
reflection from a shiny surface. All field service procedures must be followed
precisely. Only Siemens-trained field service personnel should perform
procedures related to laser assemblies.
For more safety information and laser specifications, refer to
Safety
Information, Protecting yourself from lasers
.
1.
Make sure that the analyzer is in the Standby mode.
2.
Open the optics cover. The RBC flowcell is on the left.
Perox flowcell (
1
), RBC flowcell (
2
)
3.
Disconnect the shuttle line (
1
), the sample-stream input line (
2
), and the
sheath-stream input line (
3
) from the CFM. Allow these lines to hang freely.
1
SHUTTLE NIPPLE
2
SAMPLE NIPPLE
3
SHEATH NIPPLE
4.
Locate hydraulic valve # 23. Unscrew the threaded fitting from the top left-
hand side of the valve.