Regulatory information
9-65
RETIC Method Data
The sample stability, expected ranges, and performance characteristics for the RETIC method are listed in the following Table.
Sample
Stability
Expected
Imprecision
Correlation Data
Observed
Analytical Range
Room Temp
Refrigerated
Values
Mean
SD
CV%
y=
r
Sy.x
Carryover
Range
Deviation
%RETIC
24h
72 h
0.8 to 2.1
1.2
0.16
13.3
0.82x + 0.2
0.925
0.4
0.2 to 24.9
2.6
#RETIC (10
6
/L)
40 to 79
57.8
7.7
13.3
0.79x + 7.8
0.903
13.6
CHr (pg)
24 h
72 h
24 to 36
28.9
0.2
0.7
0.96x + 4.2
0.940
1.7
RTC RBC Count
1.0%
not available or not applicable for this parameter