
Flags
6-7
Default trigger values for the three severity levels are:
+
%BLASTS 1.5% to 5.0%
++
%BLASTS > 5.0% to 10.0%
+++
%BLASTS > 10.0%
Perox Cytogram
1 LUC area
Baso Cytogram
2 BLASTS area
The %BLASTS value is intended for flagging and laboratory purposes only. The
%BLASTS value is not to be reported as a patient result.
Results Flagged
none
Possible Causes
The following list of sample-related causes may not contain all conditions that
could cause this flag.
Multiple occurrences of this flag, especially for consecutive samples, can indicate
a system problem. However, isolated instances of this flag are usually sample
related.
Sample Related
System Related
•
Acute leukemias
•
Chronic leukemias
•
Lymphomas
•
Myelofibrosis
•
Refractory anemia with
excess blasts (RAEB)
•
Neonatal samples
1.
Check the perox and baso reagents
2.
Check the delivery of SHEATH/RINSE.
3.
Check the perox hydraulics.
4.
Check baso hydraulics.
5.
Check the perox flowcell alignment.
6.
Check the laser flowcell alignment.
7.
Check the pressure and vacuum readings.
8.
Check the baso and perox gains.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...