5-50
Troubleshooting the Analyzer
Data Manager and LIS
At the host computer, how can I tell controls apart?
In the SID that is sent to the host, the first two digits after the C identify the
control as follows:
Control
Type
First
Digit
Second
Digit
CBC/DIFF
1
1, 4, or 7 = Low
2, 5,or 8 = Normal
3, 6, or 9 = High
RETIC
2
1, 3, or 5 = Low
2, 4, or 6 = High
How many 3.5" floppy disks does it take to back up 10,000 All Complete
samples?
It takes approximately 14 disks and 1.5 hours. To save time, you should do an All
Complete backup every day. During an All Complete, the system will back up
only those records that were added since the last backup. The system warns you
when it is almost full. The default warning appears at 9800 samples and can be
changed in Customer Parameters. Since the system cannot operate with more
than 10,000 records in the database, once the database nears 9000 records you
should purge the oldest 500 or 1000 records every day. This will make room for
the next day's samples. In Customer Parameters, you also have the option to set
All Complete file size to trigger purge, which will automatically purge records
when you perform the End of Day procedure.
Why does the LIS link indicator on the Sample Cont. Panel remain green
when the host-computer cable is disconnected from my PC?
The bi-directional protocol communication link sensor is time-delayed. After you
disconnect the cable, it takes from 30 to 60 seconds for the LIS link indicator to
turn red. If, however, you attempt to send information to the host computer
before this time has elapsed, the indicator will turn red immediately.
If your system is set up for monodirectional protocol, the LIS link indicator
works differently. After you turn on the host communication, the indicator stays
red until you send the first sample. Once you send the sample, the indicator turns
green and stays green even after you disconnect the host computer cable from
your PC.
No Sound When Viewing Videos
If there is no sound when you are viewing a video, adjust the sound on your
monitor. Refer to the instructions in the manual shipped with the monitor you are
using.
There is no sound with animations.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...