10-50
ADVIA Autoslide Slide Maker Stainer
Event Text
Severity
Probable Cause
Suggested Action
ASL - Smearing module
not ready
Failure
Smearing module problem.
Reset smearer. Call Siemens.
ASL - Smearing tape is
loose
Failure
Smearing film empty, broken or
incorrectly threaded.
Perform the smearing film replacement
procedure. Be sure to check smearing
wedge for cleanliness.
ASL - Stainer error
Failure
Stainer error.
Reset autoslide (see Tips # 2 & 3).
ASL - Stainer module
door is open
Failure
Stainer module door is open.
Close the stainer module door. Perform
a mechanical initialization, if necessary
ASL - Stainer not primed
Failure
Stainer is not primed.
Check message log to see if daily
startup has been done, perform a daily
startup or Stainer Prime (if daily startup
was completed) followed by a
mechanical initialization if necessary.
ASL - Stainer not ready
Failure
The stainer subsystem is not
rinsed.
Reset autoslide.
ASL - Stainer under
methanol
Failure
Stainer wells are full with methanol
Run the daily or weekly startup
procedure to drain the methanol.
ASL - Staining Error -
Stop
Failure
Stainer door was open. Run
stopped if Alarm/Stop criteria set.
Close stainer door, reset autoslide,
resume run.
ASL - Staining waste is
full
ASL - Staining Waste Full
- Stop
Failure
Staining Waste is full - stainer
enabled.
Dispose of staining waste according to
local regulations; Perform a Reset
Autoslide if necessary.
ASL - Unknown error
Failure
Autoslide reported an error
unrecognized by the ADVIA
2120/2120i system.
Call Siemens.
ASL - In Pre-smeared
slide mode
Information
An operation is rejected because
STAT slide mode is in effect.
Wait for Stat slide to be completed,
close manual slide entry door and try
again.
ASL - Autoslide aspiration
hardware unavailable on
autosampler
Warning
The autosampler hardware for the
second aspiration position is
inoperable or unavailable
If system has a dual sampler, perform
an autosampler reset. If not successful,
power the ADVIA 2120/2120i system
off/on. Call Siemens.
ASL - Autoslide is not
available to make slides
Warning
Autoslide is not ready for use.
Perform daily/weekly maintenance as
required, and try again.
ASL - Please Close
manual door
Warning
Stat door is open.
Close the manual inlet /stat door.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...