Status Line messages
Possible Cause
1
The
autosampler
performed two
index
operations, but
no aspiration.
Corrective Action
a
At the analyzer touchpad, select
Off
.
b
Wait about 1 minute and select
On
to restart the
analyzer.
c
If the problem persists, call Siemens Service for
assistance.
2
The message
sent from the
autosampler to
the analyzer
did not contain
the sample ID
sequence
number
expected for
the tube in the
current rack
position.
See
1a, b, c
Autosampler - Mechanical Constraint
To prevent the collision of autosampler components, the system did not execute
the requested action.
Autosampler - Mixer Failed
The autosampler mixer failed and may be defective.
Corrective Action
1
On the analyzer touchpad, select
Off
.
2
Wait about 1 minute and select
On
to restart the analyzer.
3
If the problem persists, call Siemens Service for assistance.
Autosampler - Mixer Lost Steps, Reset Autosampler
The autosampler mixer is not in the expected position. A reset is required.
Possible Cause
1
Car movement
problem.
Corrective Action
a
At the analyzer touchpad, select
Off
.
b
Wait about 1 minute and select
On
to restart the
analyzer.
c
If the problem persists, call Siemens Service for
assistance.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...