6-24
Flags
Flag
Triggering Criteria
CBC
CBC /
DIFF
CBC / DIFF
/ RETIC
CBC /
RETIC
RETIC
RTC-FS (FC) 1.
Difference between
mean and mode RETIC
Abs Histogram > 15%
2.
Chi-square error for the
Gaussian fit > 80,000
RTC-NO
(NO)
Noise Origin > 10% of
gated cells
RTC-L (CL)
Cells analyzed < 10,000
RTC-FL (RF) CV for RETIC Abs
Flatness histogram > 3.6
RTCSAT
(CS)
Saturation cells > 10% of
gated cells
RTC-SE (SE) Slope of negative cell
population on absorption
axis > 0.2
WBC-CE
(WC)
1.
(WBCB - WBCP)
≥
1.0
x 10³ / µL when WBCB
≥
2.0 x 10³ / µL and
≤
10.0 x 10³ / µL
2.
(WBCB - WBCP)
≥
10% of WBCB when
WBCB
≥
10.0 x 10³ / µL
3.
(WBCB-WBCP) > 20%
WBCB when WBCB
≤
2.0 x 10³ / µL
4.
WBCB count is not
flagged if the PX-NV or
PX-NO flags are set.
WBCSUB
(WS)
If WBCB is flagged but
WBCP is not, the WBCP
value is reported.
If both WBCB and WBCP
are flagged, WBCB is
reported.
If WBCB < 1000 cells/
μ
L,
WBCB is reported with an
asterisk.