7-86
Status Line messages
RETIC RBC Count Low - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The autosampler halts at the same time the stop
message occurs. The specified stop criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria. The counter associated with the Stop
Criteria is reset when the Start/Stop button is selected.
RETIC Saturation Cells - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
RETIC Saturation Cells - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The autosampler halts at the same time the stop
message occurs. The specified stop criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria. The counter associated with the Stop
Criteria is reset when the Start/Stop button is selected.
RETIC Slope Error - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
RETIC Slope Error - Stop
The number of consecutive occurrences of this condition has met the criterion
specified for a stop message. The autosampler halts at the same time the stop
message occurs. The specified stop criterion is user definable in the Customize
Menu, System Setup, Alarm/Stop Criteria. The counter associated with the Stop
Criteria is reset when the Start/Stop button is selected.
Review Mode Change to Review None
The validation mode has been modified through the Parameters menu in the
Review and Edit window.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...