
7-90
Status Line messages
Ready to Run and the green ready-to-aspirate light is
still off, follow these steps to exit the ADVIA
2120/2120i software, turn off the analyzer and then
restart.
i
Select
Shut Down ADVIA
at the Log On/Off
window.
ii
Select
Off
at the analyzer touchpad.
iii
Select
CTRL+ALT+DELETE
and log back onto the
system to restart the software.
iv
Restart the analyzer by selecting
On
at the analyzer
touchpad.
2
Faulty power
supply, fuse,
cable, selector
valve
assembly, UFC
Scrambler
Board, or Can
Bus Scrambler
Board
Call Siemens Service for assistance.
Selector Valve Timeout
The selector valve failed to return to the home position within the expected time
interval.
Possible Cause
1
Hardware failure
Corrective Action
a
Reset the analyzer by selecting
Utilities,
Exerciser, Indicators, Analyzer Reset
.
b
After about 1 minute, if the status line does not
display Ready to Run and the green ready-to-
aspirate light is still off, follow these steps to
exit the ADVIA 2120/2120i software, turn off
the analyzer and then restart.
i
Select
Shut Down ADVIA
at the Log On/Off
window.
ii
Select
Off
at the analyzer touchpad.
iii
Select
CTRL+ALT+DELETE
and log back
onto the system to restart the software.
iv
Restart the analyzer by selecting
On
at the
analyzer touchpad.
2
Faulty power
supply, fuse, cable,
selector valve
assembly, UFC
Call Siemens Service for assistance.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...