5-40
Troubleshooting the Analyzer
2.
Open the optics cover. The perox flowcell is on the left.
Perox flowcell (
1
), RBC flowcell (
2
)
3.
Disconnect the shuttle line (
1
), the sample-stream input line (
2
), and the
sheath-stream input line (
3
) from the CFM. Allow these lines to hang freely.
1
SHEATH NIPPLE
2
SAMPLE NIPPLE
3
SHUTTLE NIPPLE
4.
Locate hydraulic valve # 24, Unscrew the threaded fitting from the top right
side of the valve.
5.
Release the flowcell by loosening the release knob located on the right of the
flowcell adjustment assembly