10-48
ADVIA Autoslide Slide Maker Stainer
Event Text
Severity
Probable Cause
Suggested Action
ASL - Aspiration needle
retract timeout
Failure
Aspiration needle failed to retract
in time.
Perform Mechanical Initialization to
reset autoslide, check autosampler - if
necessary, perform an autosampler
reset or power the ADVIA 2120/2120i
system off/on.
ASL - Command Timeout
- Stop
Failure
Command timeout on autoslide.
Run stopped if Alarm/Stop criteria
set for critical error.
Perform Mechanical Initialization to
reset autoslide, Cycle power to
Autoslide if problem persists, call
Siemens.
ASL - Communications
error
Failure
Communication lost between
Analytical Module and Autoslide.
Check serial I/O cable between
instrument and autoslide, and cycle
autoslide power, perform a mechanical
initialization.
ASL - Configuration /
Parameter error
Failure
An invalid parameter was used to
configure the Autoslide.
Perform Mechanical Initialization to
reset autoslide, and report problem to
Siemens.
ASL - Critical Error - Stop
Failure
General autoslide critical error
(software error, unrecognized
version, emergency stop,
communications, bad parameter,
EEPROM error, not initialized,
sync error, sample may be diluted,
hardware error, autoslide interface
error, command timeout,
command status...). Run stopped
if Alarm/Stop criteria set.
Check message log for other errors,
correct problem, reset autoslide - see
Tip # 2.
ASL - Dispenser error
Failure
Mechanical problem on dispenser
Inspect dispenser area. Look for stuck
or broken slide. Ensure sufficient slides
are loaded to cover both slide sensors,
proceed to Tip # 2
ASL - EEPROM error
Failure
EEPROM read or write error
during initialization.
Call Siemens.
ASL - EEPROM IO error
Failure
An error occurred when reading or
updating the EEPROM.
Perform Mechanical Initialization to
reset autoslide, Cycle power to
Autoslide if problem persists, and call
Siemens.
ASL - Emergency stop
cycle
Failure
Cover may have been opened
while autoslide was busy.
Close cover, and reset autoslide. See
Tips # 2 & 3.
ASL - Hardware error
Failure
A hardware error has occurred.
Check message log for other errors,
correct problem, reset autoslide - see
Tip # 2. Call Siemens.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...