3-2
Daily Routine
Starting Each Shift
Emptying the Waste Container
BIOHAZARD WARNING
All products or objects that come in contact with human or animal body fluids should be
handled, before and after cleaning, as if capable of transmitting infectious diseases. Wear
facial protection, gloves, and protective clothing.
The operator should follow the recommendations to prevent the transmission of
infectious agents in health-care settings as recommended by the Clinical and Laboratory
Standards Institute (formerly NCCLS) in
Protection of Laboratory Workers from
Occupationally Acquired Infections; Approved Guideline - Third Edition.
2005. CLSI
Document M29-A3. This document contains complete information on user protection and
it can be used as reference material for instructions on laboratory safety.
WARNING:
Local laws and regulations protect the environment and encourage resource
conservation by regulating the disposal of hazardous wastes. Because some of the wastes
generated by the analyzer may be classified as hazardous waste, you must be familiar
with the applicable hazardous waste handling and disposal laws and regulations in your
area.
All analyzer waste is stored in an 11-liter waste container. When the fluid level in
the container reaches the “Maximum Level” line (approximately 8 liters), an
error message appears on the monitor and an audible alarm is sounded. The
system will not aspirate any more samples until the waste container is emptied.
Manual Waste Removal
CAUTION:
Make sure all cycles are completed before disconnecting the waste container.
The waste container
must
be connected to the system when any system cycles are in
progress, including the system startup cycle. If the waste container is not connected, fluid
will back up into the UFC and damage the system.
1.
Make sure that the analyzer is not sampling.
2.
If a message requesting a rinse or a wash
displays, select
Cancel
.
3.
Disconnect the waste line (
1
) and the
vacuum line (
2
). To do this, select the
buttons (
3
) on the quick-release connectors
as you pull the lines straight up.
4.
Disconnect the level switch sensor
connector (
4
) by selecting its button.
5.
Replace the full container with an empty one and connect the waste and
vacuum lines and level sensor to the new container.
CAUTION:
Do not open the waste container cap! If the cap is loosened or not replaced
correctly, sufficient operating vacuum cannot be reached.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...