Status Line messages
7-71
PEROX 3 Reagent Expired
Present date is beyond the PEROX 3 reagent expiration date, as determined by
encoded reagent container barcode label or as calculated from installation date
and open container stability.
Corrective Action
Replenish expired reagents and update the Reagent Installation window. Go to
Logs, Reagent Log, Reagent Installation.
PEROX 3 Reagent Low - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message, indicating that the reagent level is low. The
specified alarm criterion is user definable in the Customize Menu, System Setup,
Alarm/Stop Criteria, Reagent Conditions.
Corrective Action
The alarm counter is automatically reset after reagents are replenished and the
Reagent Installation window is updated. Go to Logs, Reagent Log, Reagent
Installation.
PEROX Irregular Flow Rate - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
PEROX Irregular Flow Rate - Stop
The number of cycles specified for a stop message. The specified stop criterion is
user definable in the Customize Menu, System Setup, Alarm/Stop Criteria. The
counter associated with the Stop Criteria is reset when the Start/Stop button is
selected.
PEROX Power Low - Alarm
The number of consecutive occurrences of this condition has met the criterion
specified for an alarm message. The specified alarm criterion is user definable in
the Customize Menu, System Setup, Alarm/Stop Criteria. The counter associated
with the Alarm Criteria is automatically reset.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...