
Status Line messages
7-83
Possible Cause
1
RBC Sheath
syringe pump
not moving or
not in home
position.
Corrective Action
a
Reset the analyzer by selecting
Utilities,
Exerciser, Indicators, Analyzer Reset
.
b
After about 1 minute, if the status line
does not display Ready to Run and the
green ready-to-aspirate light is still off,
follow these steps to exit the ADVIA
2120/2120i software, turn off the analyzer
and then restart.
i
Select
Shut Down ADVIA
at the Log
On/Off window.
ii
Select
Off
at the analyzer touchpad.
iii
Select
CTRL+ALT+DELETE
and log
back onto the system to restart the
software.
iv
Restart the analyzer by selecting
On
at
the analyzer touchpad.
2
RBC Sheath
syringe
incorrectly
assembled.
Follow the steps in Replacing the Syringe
Plungers.
3
Faulty power
supply, fuse,
cable, pump
assembly, Can
Bus Scrambler
Board or Dual
Servo Pump
Control Board.
Call Siemens Service for assistance.
Ready to Run
The system is ready to accept the aspiration of a sample or the start of a hydraulic
cycle. The green ready-to-aspirate light is lit.
Results Received for Tube Not in
Pending File Sid#: xxx Seq#: A set of results has been received for a sample that
is not in the Pending file. The set of results has been refused and deleted.
Corrective Action
Check pending list before run.
Содержание ADVIA 2120
Страница 1: ...ADVIA 2120 2120i Hematology Systems Operator s Guide 067D0157 01 Rev 201 0 ...
Страница 30: ...1 26 Welcome to the ADVIA 2120 2120i Hematology System ...
Страница 46: ...3 12 Daily Routine ...
Страница 171: ...7 2 Status Line messages T 93 U 94 V 96 W 98 X 99 Y 99 Z 99 ...
Страница 269: ...7 100 Status Line messages ...
Страница 341: ...8 72 Methods ...
Страница 477: ...B 14 Warnings and Safety Information ...